Blog

Testing Report of Ozone Analyzer UV-2000S

O3tech 2014-6-22 14:23:02

Testing Report of Ozone Analyzer UV-2000S

O3 Tech (H.K) Limited

BLDG C#, MEICHENG TECH PARK, CHANG ZHEN, GONGMING, GUANGMING NEW DISTRICT,
SHENZHEN, CHINA 518000
Tel: 86.755.89969024
http://www.o3tech.cn; info@o3tech.cn

Testing  Report






14-Apr-16
Ozone AnalyzerUltraviolet Absorption. UV-2000S

Cooling method: air cooling




Temperature: 25  





 

ModelOzone Tubes QTY  VoltageBottle Oxygen O2 Flow/minOzone Concentration Mg/LOzone Output grams/H
SP001G    1  220V/HZ    99%    2     27.18    3.2616




    1     34.78    2.0868




    1/4     45.56     0.6834




    1/8     43.82       0.32865




    1/16     42.68        0.16005

Ozone output calculating formula: 






Oxygen flow (LPM) ×60 Min(1 hour)× ozone concentration (MG/L)÷1000=Ozone output (Grams/Hour)